Used by technical divers, the Nuvair Pro Trimix Alarm measures helium (He) and oxygen (O2) levels in mixed gases. Self-calibrating with audible and visual alarms. With an optional relay, the Pro Trimix Alarm can shut down a compressor at user-defined set points.
The Nuvair Pro Trimix Alarm analyzer measures helium (He) and oxygen (O2) levels in mixed breathing gases stored in gas cylinders or from the continuous flow of gas from a compressor. The Analyzer is compatible with outdoor and marine environments, is self-calibrating and includes user-defined audible and visual set point alarms. If the alarm sounds, the analog output can activate an optional relay to shut down equipment automatically, sound a remote alarm, or activates an emergency backup system. Range: 0-100% with display accuracy of ±2%. If the alarm sounds, the analog output can activate an optional relay that shuts down compressors automatically, sounds a remote alarm, or activates an emergency backup system.
PLEASE NOTE: Never expose gas sensors to pressure or you may cause damage and/or false readings. Damaged sensors will not provide accurate gas analysis. Inaccurate gas analysis can lead to serious personal injury or death. Most gas analyzers can be used to analyze a regulated gas sample flow, the contents of a gas cylinder, or the flow from a regulator. The flow rate of gas must equal 1-5 L/min. To produce this flow, a Flow Restrictor and Regulator may be required.
Specifications:
SENSORS
|
Pro He
|
PRO O2
|
|
Flow
Rate
|
0.5-1 L/min
|
0.5-2 L/min
|
|
Resolution
|
0.10%
|
0.10%
|
|
Linearity
/ Repeatability
|
±2% over full scale (Linearity)
|
± 1 % volume O2 @ 100% O2, applied for 5 min (Repeatability)
|
|
Accuracy
|
±2% over full scale
|
±1%
over full scale if sensor well calibrated
and @ constant temperature, pressure and gas flow
|
|
|
Sensor
Type
|
Thermal conductive
|
Electrochemical
|
|
Expected
Sensor Life
|
>24 months
|
36 months
|
|
Range
|
0.0-100.0% helium in air or nitrogen or oxygen
|
0-100% oxygen
|
|
Response
Time
|
<10 seconds for 90% response
|
<6 seconds for 90% of final value
|
|
Alarms
|
(2) User-programmable audible and visual alarms
|
(2) User-programmable audible and visual alarms
|
|
Operating
Temperature
|
41-104°F (5-40°C)
|
32-104°F (0-40°C)
|
|
Storage
Temperature
|
5-122°F (15 to 50°C)
|
32-122°F (0-50°C)
|
|
Operating
Humidity
|
0-90% rh, non-condensing
|
0-99% rh, non-condensing
|
|
Power
|
Rechargeable lithium battery with 110/230 V charger
|
|
Dimensions
& Weight
|
3.9 x 8.7 x 7.5 in (9.9 x 22.1 x 19.1 cm)
|
|
2.75 lb (1.25 kg)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Features
- On/Off Switch Display Range of 0.0-100%
- Two (2) custom audible and visual alarms for each gas
- 4-20 mA analog output for external devices
- Capable of shutting down compressor at set alarm with optional relay cable
|
|
|
|
|
|
|